Showing posts with label Leadless Pacing. Show all posts
Showing posts with label Leadless Pacing. Show all posts

Thursday, February 1, 2018

Normal Impulse Conduction

Sequence of Cardiac Electrical Activation

conduction velocities in the heart
The action potentials generated by the SA node spread throughout the atria primarily by cell-to-cell conduction at a velocity of about 0.5 m/sec. There is some functional evidence for the existence of specialized conducting pathways within the atria (termed internodal tracts), although this is controversial. As the wave of action potentials depolarizes the atrial muscle, the cardiomyocytes contract by a process termed excitation-contraction coupling.
Normally, the only pathway available for action potentials to enter the ventricles is through a specialized region of cells (atrioventricular node, or AV node) located in the inferior-posterior region of the interatrial septum. The AV node is a highly specialized conducting tissue (cardiac, not neural in origin) that slows the impulse conduction considerably (to about 0.05 m/sec) thereby allowing sufficient time for complete atrial depolarization and contraction (systole) prior to ventricular depolarization and contraction.

heart depolarization times
The impulses then enter the base of the ventricle at the Bundle of His and then follow the left and right bundle branches along the interventricular septum.  These specialized fibers conduct the impulses at a very rapid velocity (about 2 m/sec).  The bundle branches then divide into an extensive system of Purkinje fibers that conduct the impulses at high velocity (about 4 m/sec) throughout the ventricles. This results in rapid depolarization of ventricular myocytes throughout both ventricles.

The conduction system within the heart is very important because it permits a rapid and organized depolarization of ventricular myocytes that is necessary for the efficient generation of pressure during systole. The time (in seconds) to activate the different regions of the heart are shown in the figure to the right. Atrial activation is complete within about 0.09 sec (90 msec) following SA nodal firing. After a delay at the AV node, the septum becomes activated (0.16 sec). All the ventricular mass is activated by about 0.23 sec.

Regulation of Conduction


The conduction of electrical impulses throughout the heart, and particularly in the specialized conduction system, is influenced by autonomic nerve activity. This autonomic control is most apparent at the AV node. Sympathetic activation increases conduction velocity in the AV node by increasing the rate of depolarization (increasing slope of phase 0) of the action potentials. This leads to more rapid depolarization of adjacent cells, which leads to a more rapid conduction of action potentials (positive dromotropy). Sympathetic activation of the AV node reduces the normal delay of conduction through the AV node, thereby reducing the time between atrial and ventricular contraction. The increase in AV nodal conduction velocity can be seen as a decrease in the P-R interval of the electrocardiogram.


sympathetic and vagal effcts on atrioventricular node action potentials

Sympathetic nerves exert their actions on the AV node by releasing the neurotransmitter norepinephrine that binds to beta-adrenoceptors, leading to an increase in intracellular cAMP. Therefore, drugs that block beta-adrenoceptors (beta-blockers) decrease conduction velocity and can produce AV block.


Parasympathetic (vagal) activation decreases conduction velocity (negative dromotropy) at the AV node by decreasing the slope of phase 0 of the nodal action potentials. This leads to slower depolarization of adjacent cells, and reduced velocity of conduction. Acetylcholine, released by the vagus nerve, binds to cardiac muscarinic receptors, which decreases intracellular cAMP. Excessive vagal activation can produce AV block. Drugs such as digitalis, which increase vagal activity to the heart, are sometimes used to reduce AV nodal conduction in patients that have atrial flutter or fibrillation. These atrial arrhythmias lead to excessive ventricular rate (tachycardia) that can be suppressed by partially blocking impulses being conducted through the AV node.
Phase 0 of action potentials at the AV node is not dependent on fast sodium channels as in non-nodal tissue, but instead is generated by the entry of calcium into the cell through slow-inward, L-type calcium channels. Blocking these channels with a calcium-channel blocker such as verapamil or diltiazem reduces the conduction velocity of impulses through the AV node and can produce AV block.
Because conduction velocity depends on the rate of tissue depolarization, which is related to the slope of phase 0 of the action potential, conditions (or drugs) that alter phase 0 will affect conduction velocity.  For example, conduction can be altered by changes in membrane potential, which can occur during myocardial ischemia and hypoxia. In non-nodal cardiac tissue, cellular hypoxia leads to membrane depolarization, inhibition of fast Na+ channels, a decrease in the slope of phase 0, and a decrease in action potential amplitude. These membrane changes result in a decrease in speed by which action potentials are conducted within the heart. This can have a number of consequences. First, activation of the heart will be delayed, and in some cases, the sequence of activation will be altered. This can seriously impair ventricular pressure development. Second, damage to the conducting system can precipitate tachyarrhythmias by reentry mechanisms.

Monday, January 8, 2018

Leadless Pacing

Leadless pacing available for selected patients


Early pacing devices offered single-chamber, fixed-rate ventricular pacing for life-threatening conduction system disease. Advances in generator and lead technology and the results of clinical trials over the past 60 years have expanded the indications for device therapy. As a result, more individuals are receiving device therapy; approximately 190,000 pacemakers are implanted every year in the United States.

Over time, the patient population receiving pacing therapy has become older and more complex. The "weak link" in device therapy has been the leads. While transvenous leads typically have lower thresholds and better longevity than epicardial leads, they are associated with increased morbidity and mortality.

Complications at implant include bleeding, vascular damage, cardiac perforation, pneumothorax and dislodgment. Potential long-term concerns include lead fracture, malfunction, venous obstruction, tricuspid valve regurgitation and the risks associated with lead extraction. Transvenous leads are contraindicated in the presence of right-to-left shunt and in some patients with congenital heart disease.
The concept of leadless pacing was first proposed in 1970 by J. William Spickler, Ph.D., and colleagues, but it is only recently that leadless devices have become available. Currently available are the Nanostim leadless pacemaker (St. Jude Medical, St. Paul, Minnesota) and the Micra transcatheter pacing system (Medtronic, Minneapolis); both are dime-sized capsules that are implanted directly into the right ventricular apex.

The Nanostim uses active fixation, while the Micra has a tined fixation mechanism to secure the device to the right ventricular endocardial surface. Both devices are capable of VVIR pacing and have estimated battery longevity between 7 and 10 years. When the battery is depleted, a new device can be implanted and the existing device left in place.
These devices are contraindicated in individuals who require dual-chamber pacing or who have demonstrable pacemaker syndrome. Anticoagulation is not required after implant placement. Current devices are not MRI compatible.

Leadless pacemakers are contraindicated in patients with implantable cardioverter-defibrillators, as high-voltage shocks could damage the pacemaker, and the effect of the pacemaker on shock effectiveness is unknown.

Leadless devices should be avoided in individuals with elevated right ventricular pressures because of higher theoretical risk of embolization. The presence of mechanical tricuspid valves or inferior vena cava filters also precludes the use of leadless pacemakers. Successful device retrieval has been accomplished in animal studies.

Sunday, January 7, 2018

Sudden death in young people: Heart problems often blamed



Sudden death in young people is rare, but those at risk can take precautions. Find out more about the risk factors, causes and treatments.

Sudden death in people younger than 35, often due to undiscovered heart defects or overlooked heart abnormalities, is rare. When these sudden deaths occur, it's often during physical activity, such as playing a sport, and more often occurs in males than in females.
Millions of elementary, high school and college athletes compete yearly without incident. If you or your child is at risk of sudden death, talk to your doctor about precautions you can take.

How common is sudden cardiac death in young people?


Most deaths due to cardiac arrest are in older adults, particularly those with coronary artery disease. Cardiac arrest is the leading cause of death in young athletes, but the incidence of it is unclear. Perhaps 1 in every 50,000 sudden cardiac deaths a year occurs in young athletes.

What can cause sudden cardiac death in young people?


The causes of sudden cardiac death in young people vary. Most often, death is due to a heart abnormality.
For a variety of reasons, something causes the heart to beat out of control. This abnormal heart rhythm is known as ventricular fibrillation.
Some specific causes of sudden cardiac death in young people include:
  • Hypertrophic cardiomyopathy (HCM). In this usually inherited condition, the walls of the heart muscle thicken. The thickened muscle can disrupt the heart's electrical system, leading to fast or irregular heartbeats (arrhythmias), which can lead to sudden cardiac death.
  • Hypertrophic cardiomyopathy, although not usually fatal, is the most common cause of heart-related sudden death in people under 30. It's the most common identifiable cause of sudden death in athletes. HCM often goes undetected.
  • Coronary artery abnormalities. Sometimes people are born with heart arteries (coronary arteries) that are connected abnormally. The arteries can become compressed during exercise and not provide proper blood flow to the heart.
  • Long QT syndrome. This inherited heart rhythm disorder can cause fast, chaotic heartbeats, often leading to fainting. Young people with long QT syndrome have an increased risk of sudden death.

Other causes of sudden cardiac death in young people include structural abnormalities of the heart, such as undetected heart disease that was present at birth (congenital) and heart muscle abnormalities.
Other causes include inflammation of the heart muscle, which can be caused by viruses and other illnesses. Besides long QT syndrome, other abnormalities of the heart's electrical system, such as Brugada syndrome, can cause sudden death.
Commotio cordis, another rare cause of sudden cardiac death that can occur in anyone, occurs as the result of a blunt blow to the chest, such as being hit by a hockey puck or another player. The blow to the chest can trigger ventricular fibrillation if the blow strikes at exactly the wrong time in the heart's electrical cycle.


Are there symptoms or red flags parents, coaches and others should be on the lookout for that signal a young person is at high risk of sudden cardiac death?


Many times these deaths occur with no warning, indications to watch for include:
  • Unexplained fainting (syncope). If this occurs during physical activity, it could be a sign that there's a problem with your heart.
  • Family history of sudden cardiac death. The other major warning sign is a family history of unexplained deaths before the age of 50. If this has occurred in your family, talk with your doctor about screening options.
Shortness of breath or chest pain could indicate that you're at risk of sudden cardiac death. They could also indicate other health problems in young people, such as asthma.


Can sudden death in young people be prevented?

Sometimes. If you're at high risk of sudden cardiac death, your doctor will usually suggest that you avoid competitive sports. Depending on your underlying condition, medical or surgical treatments might be appropriate to reduce your risk of sudden death.
Another option for some, such as those with hypertrophic cardiomyopathy, is an implantable cardioverter-defibrillator (ICD). This pager-sized device implanted in your chest like a pacemaker continuously monitors your heartbeat. If a life-threatening arrhythmia occurs, the ICD delivers electrical shocks to restore a normal heart rhythm.


Who should be screened for sudden death risk factors?

There's debate in the medical community about screening young athletes to attempt to identify those at high risk of sudden death.
Some countries such as Italy screen young people with an electrocardiogram (ECG or EKG), which records the electrical signals in the heart. However, this type of screening is expensive and can produce false-positive results — indications that an abnormality or disease is present when it isn't — which can cause unnecessary worry and additional tests.
It's not clear that routine exams given before athletes are cleared to play competitive sports can prevent sudden cardiac death. However, they might help identify some who are at increased risk.
For anyone with a family history or risk factors for conditions that cause sudden cardiac death, further screening is recommended. Repeat screening of family members is recommended over time, even if the first heart evaluation was normal.


Should young people with a heart defect avoid physical activity?

If you're at risk of sudden cardiac death, talk to your doctor about physical activity. Whether you can participate in exercise or sports depends on your condition.
For some disorders, such as hypertrophic cardiomyopathy, it's often recommended that you avoid most competitive sports and that if you have an ICD, you should avoid impact sports. But this doesn't mean you need to avoid exercise. Talk to your doctor about restrictions on your activity.



Implantable Cardioverter-defibrillators (ICDs)



Overview

An implantable cardioverter-defibrillator (ICD) — a pager-sized device — is placed in your chest to reduce your risk of dying if the lower chambers of your heart (ventricles) go into a dangerous rhythm and stop beating effectively (cardiac arrest).
You might need an ICD if you have a dangerously fast heartbeat (ventricular tachycardia) or a chaotic heartbeat that keeps your heart from supplying enough blood to the rest of your body (ventricular fibrillation).
ICDs detect and stop abnormal heartbeats (arrhythmias). The device continuously monitors your heartbeat and delivers electrical pulses to restore a normal heart rhythm when necessary.
An ICD differs from a pacemaker — another implantable device used to help control abnormal heart rhythms.

Why it's done

You've likely seen TV shows in which hospital workers "shock" an unconscious person out of cardiac arrest with electrified paddles. An ICD does the same thing only internally and automatically when it detects an abnormal heart rhythm.
An ICD is surgically placed under your skin, usually below your left collarbone. One or more flexible, insulated wires (leads) run from the ICD through your veins to your heart.
Because the ICD constantly monitors for abnormal heart rhythms and instantly tries to correct them, it helps treat cardiac arrest, even when you are far from the nearest hospital.

How an ICD works

When you have a rapid heartbeat, the wires from your heart to the device transmit signals to the ICD, which sends electrical pulses to regulate your heartbeat. Depending on the problem with your heartbeat, your ICD could be programmed for the following therapies:
  • Low-energy pacing therapy. You may feel nothing or a painless fluttering in your chest when your ICD responds to mild disruptions in your heartbeat.
  • Cardioversion therapy. A higher energy shock is delivered for a more serious heart rhythm problem. It may feel as if you're being thumped in the chest.
  • Defibrillation therapy. This is the strongest form of electrical therapy used to restore a normal heartbeat. During this therapy, it may feel as if you're being kicked in the chest, and it might knock you off your feet.
    The pain from this therapy usually lasts only a second. There should be no discomfort after the shock ends.
Usually, only one shock is needed to restore a normal heartbeat. Sometimes, however, you might have two or more shocks during a 24-hour period. Frequent shocks in a short time period are known as ICD storms. If you have ICD storms, you should seek emergency care to see if your ICD is working properly or if you have a problem that's making your heart beat abnormally.
If necessary, the ICD can be adjusted to reduce the number and frequency of shocks. You may need additional medications to make your heart beat regularly and decrease the chance of an ICD storm.
An ICD can also be programmed to perform other functions, which include:
  • Antitachycardia (tak-ih-KAHR-dee-uh). If you have an unusually fast heart rate, the ICD delivers painless, low-energy impulses that pace or stimulate the heart to beat normally. This can prevent the need for cardioversion or defibrillation.
  • Pacemaker. Most modern ICDs can also function as a pacemaker, delivering low-energy impulses that stimulate the heart to beat normally.
  • Recording heart activity. The ICD records information about variations in your heart's electrical activity and rhythm. This information helps your doctor evaluate your heart rhythm problem and, if necessary, reprogram your ICD.

Subcutaneous ICD

A subcutaneous ICD (S-ICD), a newer type of ICD, is available at some surgical centers. An S-ICD is implanted under the skin at the side of the chest below the armpit. It's attached to an electrode that runs along your breastbone. You may be a candidate for this device if you have structural defects in your heart that prevent attaching wires to the heart your blood vessels, or if you have other reasons for wanting to avoid traditional ICDs.
Implanting a subcutaneous ICD is less invasive than an ICD that attaches to the heart, but the device is larger in size than an ICD.

Who needs an ICD

You're a candidate for an ICD if you've had sustained ventricular tachycardia, survived a cardiac arrest, or fainted from a ventricular arrhythmia. You might also benefit from an ICD if you have:
  • A history of coronary artery disease and heart attack that has weakened your heart.
  • A heart condition that involves abnormal heart muscle, such as enlarged (dilated cardiomyopathy) or thickened (hypertrophic cardiomyopathy) heart muscle.
  • An inherited heart defect that makes your heart beat abnormally. These include long QT syndrome, which can cause ventricular fibrillation and death even in young people with no signs or symptoms of heart problems.
  • Having other rare conditions such as Brugada syndrome and arrhythmogenic right ventricular dysplasia also may mean you need an ICD.

Risks

Risks associated with ICD implantation are uncommon but may include:
  • Infection at the implant site
  • Allergic reaction to the medications used during the procedure
  • Swelling, bleeding or bruising where your ICD was implanted
  • Damage to the vein where your ICD leads are placed
  • Bleeding around your heart, which can be life-threatening
  • Blood leaking through the heart valve where the ICD lead is placed
  • Collapsed lung (pneumothorax)

How you prepare

To determine whether you need an ICD, your doctor might perform a variety of diagnostic tests, which may include:
  • Electrocardiography (ECG). In this noninvasive test, sensor pads with wires attached (electrodes) are placed on your chest and sometimes limbs to measure your heart's electrical impulses. Your heart's beating pattern can offer clues to the type of irregular heartbeat you have.
  • Echocardiography. This noninvasive test uses harmless sound waves that allow your doctor to see your heart without making an incision. During the procedure, a small, plastic instrument called a transducer is placed on your chest.
    It collects reflected sound waves (echoes) from your heart and transmits them to a machine that uses the sound wave patterns to compose images of your beating heart on a monitor. These images show how well your heart is functioning, and recorded images allow your doctor to measure the size and thickness of your heart muscle.
  • Holter monitoring. Also known as an ambulatory electrocardiogram monitor, a Holter monitor records your heart rhythm for 24 hours. Wires from electrodes on your chest go to a battery-operated recording device carried in your pocket or worn on a belt or shoulder strap.
    While wearing the monitor, you'll keep a diary of your activities and symptoms. Your doctor will compare the diary with the electrical recordings and try to figure out the cause of your symptoms.
  • Event recorder. Your doctor might ask you to wear a pager-sized device that records your heart activity for more than 24 hours. Unlike a Holter monitor, it doesn't operate continuously — you turn it on when you feel your heart is beating abnormally.
  • Electrophysiology study (EPS). Electrodes are guided through blood vessels to your heart and used to test the function of your heart's electrical system. This can identify whether you have or might develop heart rhythm problems.
It's likely you'll be asked not to eat or drink for at least eight hours before your surgery. Talk to your doctor about the medications you take and whether you should continue to take them before your procedure to implant an ICD.

What you can expect

During the procedure

Usually, the procedure to implant an ICD can be performed with numbing medication and a sedative that relaxes you but allows you to remain aware of your surroundings. In some cases, general anesthesia, which puts you to sleep, may be used.
The procedure usually takes a few hours. During surgery, one or more flexible, insulated wires (leads) are inserted into veins near your collarbone and guided, with the help of X-ray images, to your heart. The ends of the leads are secured to your heart, while the other ends are attached to the generator, which is usually implanted under the skin beneath your collarbone.
Once the ICD is in place, your doctor will test it and program it for your heart rhythm problem. Testing the ICD might require speeding up your heart and then shocking it back into normal rhythm.

After the procedure

You'll stay in the hospital one or two days, and the ICD might be tested once more before you're discharged. Additional testing of your ICD usually doesn't require surgery.

Treating pain after your procedure

After surgery, you may have some pain in the incision area, which can remain swollen and tender for a few days or weeks. Your doctor might prescribe pain medication. As your pain lessens, you can take nonaspirin pain relievers, such as acetaminophen (Tylenol, others) or ibuprofen (Advil, Motrin IB, others).
Unless your doctor instructs you to do so, don't take pain medication containing aspirin because it can increase bleeding risk.
When you're released from the hospital, you'll need to arrange for a ride home because you won't be able to drive for at least a week.

Results

ICDs have become standard treatment for anyone who has survived cardiac arrest, and they're increasingly used in people at high risk of sudden cardiac arrest. An ICD lowers your risk of sudden death from cardiac arrest more than medication alone.
Although the electrical shocks can be unsettling, there is evidence that the ICD is effectively treating your heart rhythm problem and protecting you from sudden death. Talk to your doctor about how to best care for your ICD.
After the procedure, you'll need to take some precautions to avoid injuries and make sure your ICD works properly.

Short-term precautions

You'll likely be able to return to normal activities such as exercise, work and sex soon after you recover from surgery. Follow your doctor's instructions. For four weeks after surgery, your doctor might ask you to refrain from:
  • Vigorous above-the-shoulder activities or exercises, including golf, tennis, swimming, bicycling, bowling or vacuuming
  • Lifting anything weighing more than 5 pounds
  • Playing contact sports
  • Strenuous exercise programs

Long-term precautions

Problems with your ICD due to electrical interference are rare. Still, take precautions with the following:
  • Cellular phones and other mobile devices. It's safe to talk on a cellphone, but avoid placing your cellphone within 6 inches (about 15 centimeters) of your ICD implantation site when the phone is turned on. Although unlikely, your ICD could mistake a cellphone's signal for a heartbeat and slow your heartbeat, causing symptoms such as sudden fatigue.
  • Security systems. After surgery, you'll receive a card that says you have an ICD. Show your card to airport personnel because the ICD may set off airport security alarms.
    Also, hand-held metal detectors often contain a magnet that can interfere with your ICD. Limit scanning with a hand-held detector to less than 30 seconds over the site of your ICD or make a request for a manual search.
  • Medical equipment. Let doctors, medical technicians and dentists you see know you have an ICD. Some procedures, such as magnetic resonance imaging (MRI), magnetic resonance angiography (MRA), and radiofrequency or microwave ablation are not recommended if you have an ICD.
  • Power generators. Stand at least 2 feet (0.6 meters) from welding equipment, high-voltage transformers or motor-generator systems. If you work around such equipment, your doctor can arrange a test in your workplace to see if the equipment affects your ICD.
  • MP3 player headphones. Although the player itself poses little risk, the headphones may be a problem. Most contain a magnetic substance and can interfere with your ICD. Keep your headphones at least 6 inches (about 15 centimeters) from your ICD.
  • Magnets. These might affect your ICD, so it's a good idea to keep magnets at least 6 inches (15 centimeters) from your ICD site.
Devices that pose little or no risk to your ICD include microwave ovens, televisions and remote controls, AM/FM radios, toasters, electric blankets, electric shavers and electric drills, computers, scanners, printers, and GPS devices.

Driving restrictions

If you have an ICD to treat ventricular arrhythmia, driving a vehicle presents a challenge. The combination of arrhythmia and shocks from your ICD can cause fainting, which would be dangerous while you're driving.
The American Heart Association's guidelines advise avoiding driving for one week after ICD placement, but talk to your doctor for specific recommendations. The guidelines discourage driving during the first six months after your procedure if your ICD was implanted due to a previous cardiac arrest or ventricular arrhythmia.
If you have no shocks during this period, you'll likely be able to drive again. But if you then have a shock, with or without fainting, tell your doctor and follow his or her recommendations. In most cases, you'll need to stop driving until you've been shock-free for another six months.
If you have an ICD but have no history of life-threatening arrhythmias, you can usually resume driving within a week after your procedure if you've had no shocks. Discuss your situation with your doctor.
You usually can't get a commercial driver's license if you have an ICD.

Battery life

The lithium battery in your ICD can last up to seven years. The battery will be checked during regular checkups, which should occur every three to six months. When the battery is nearly out of power, your old shock generator is replaced with a new one during a minor outpatient procedure.

ICDs and end-of-life issues

If you have an ICD and become terminally ill, your ICD could deliver unnecessary shocks. It's easy to turn off your ICD, and turning it off may prevent unnecessary suffering.
Talk to your doctor about your wishes. Also talk to family members or another person designated to make medical decisions for you about what you'd like to do in end-of-life care situations.



History of Heart and Electricity ( A brief history of cardiac pacing )






A brief history of cardiac pacing



Introduction

The history of cardiac pacing therapy must be viewed within the broader framework of electro-diagnosis and electro-therapy. Moreover it can be observed that the development of electro-therapy usually preceded the understanding of what was actually occurring within the heart.
Electro-therapy has a simple core concept: the use of an outside source of electricity to stimulate human tissue in various ways to produce a beneficial therapeutic effect. This has shown a prolonged, halting development through the ages, sometimes being looked upon as mysterious magic produced by complex machines.
Over the last fifty years or so, electro-therapy has shown a very rapid, almost explosive, development with many innovators contributing to a whole series of “firsts”. This was the consequence of a remarkable co-operation among surgeons, physicians, engineers, chemists, businessmen and patients. The field of paediatric open heart surgery gave a major impetus to the development of pacemakers since heart block often accompanied impeccably performed intra-cardiac repairs of congenital defects.
Most of the scientists and physicians involved in electro-therapy faced significant criticism and sometimes even derision by the contemporary scientific community. Yet the speciality moved steadily on gaining medical respectability and now helps countless patients all over the world.


Antiquity


Hippocrates (460 – 375 BC) (Fig. 1)


“Those who suffer from frequent and strong faints without any manifest cause die suddenly”

Aristotle (384 – 322 BC) (Fig. 2)


Aristotle saw the heart as “the source of all movement, since the heart links the soul with the organs of life”. Remarkably accurite as a description of cardiovascular physiology!


The pulse

In ancient China (280 BC), Wang Shu-he wrote 10 books about the pulse. The Greeks called the pulse “sphygmos”, and sphygmology thus deals with the theory of pulse. Galen, in Roman times, interpreted the various types of pulses according to current thought, that each organ in each disease has its own form of pulse .



Arabian Science age


Avicenna (ca. 980–1037) wrote the major medical textbook, The Canon of Medicine. ibn al-Nafis (1213–1288) wrote an influential book on medicine; it is believed to have replaced Avicenna's Canon in the Islamic world. He wrote commentaries on Galen and Avicenna's works. One of these commentaries, discovered in 1924, described the circulation of blood through the lungs .


Ancient Roman physicians

These treated patients suffering from pain and acute gout with electric rays (Fig. 3) and other electrically-charged sea creatures. Electric rays are cartilaginous fish with two large, kidney-shaped electric organs on either side of their head. These organs are capable of generating strong electric shocks, which are administered at will to deter predators or capture prey.

1580

Geronimo Mercuriale (Fig. 4) formulated the concept of syncope and demonstrated its connection with a slow pulse rate: “Ubi pulsus sit rarus semper expectanda est syncope”.

1600

William Harvey restarted an arrested pigeon's heart by a simple flick of the finger. In 1628 he described the circulation (Fig. 5).

1713

While the anatomy and physiology of the heart were being studied, others analysed the peripheral pulse, which was the mechanical expression of cardiac activity. In 1713, Micheal Bernhard Valentini used pulse diagrams and pulse theory in the general practice of medicine (Fig. 6).


Early cardiac electrotherapy

1640's

Publications speculating on the bio-electric nature of the cardiovascular system appeared.

1774

The first reference to external electrical stimulation of the heart in the Registers of the Royal Human Society of London. The physician was Squires and the patient was a young girl.

1775

The Danish physicist Nickolev Abildgaard conducted the first studies on the effects of electrical energy when applied to the body. He placed electrodes on the sides of a hen's head and applied an electric discharge which caused it to fall dead. Application of electrodes over various parts of the hen's body failed to reanimate the bird, until they were placed across the chest. In this position they presumably defibrillated the heart after which the hen staggered onto its feet and walked away (Fig. 7).

1791

Luigi Galvani (Fig.8), an Italian physician and natural scientist, announced that electricity was inherent in organic tissue. He published the experimental findings of electrical phenomena in frog muscles and frog hearts making a fundamental contribution to modern cardiac electrophysiology.
There was general agreement that electricity had a pronounced effect on the heart (Fig.(Fig.9,9, ,1010).

1797

Alexander von Humboldt found a dead bird in his garden and placed a blade of zinc in the beak and a shaft of silver into the rectum. An electric shock caused the bird to flap its wings and attempt to walk. He also tried the experiment on himself with unpleasant consequences.

The 19th century

Rudimentary forms of electrical stimulation were used by physicians sporadically to treat cardiac disease in numerous ways without any standardisation. The crude technology was however far ahead of the understanding of heart disease and a very wide range of effects were documented.

1800

The Italian physicist, Alessandro Volta (Fig. 11) showed that current electricity could be produced by the contact of dissimilar metals and devised the first electric battery for low-voltage high-current stimulation, the “voltaic pile” (Fig. 12). For the first time, electricity could be produced by means other than through electrostatic machines. He also gave his name to one of the basic physical units of electricity: the Volt, unit of electromotive force.

1800 – 1802

Marie Francois Xavier Bichat (Fig. 13) and Nysten reported experiments on decapitated humans in whom they were able to make hearts beat again using electric current. They had no shortage of experimental material during the French Revolution.

1804

Aldini (1762 - 1834) (Fig. 14), described the alleviation of cardiac syncope through "galvanic energy" utilising animal and cadaver studies.

1855

Rudolph Albert von Kollicker published work on the “action currents” of the heart and showed that a definite electric current was produced with each beat of a frog's heart.

1872

Duchenne de Boulogne (1806 - 1875), successfully resuscitated a child who had drowned by attaching one electrode to a leg while rhythmically tapping the precordium with another electrode.

1882

A golden opportunity for clinical and scientific experimentation arose in 1882. A 46-year old female patient arrived in the clinic of Hugo Von Ziemssen (Fig. 15). She was an unskilled labourer called Catharina Serafin from Upper Silesia in Prussia (Fig. 16). A chest tumour had been excised together with the left anterior part of her thoracic wall thus exposing her heart, which could be seen through a thin layer of skin. Von Ziemssen stimulated her heart using electric current and could change her heart rate at will. The recordings (Fig. 17) clearly show that ventricular activity was being produced by electrical impulses applied to the cardiac surface: extremely interesting but potentially fatal investigations!

Late 1800's

The English doctor John Mac William (Fig. 18) collected and analysed all the scattered data available at that time and laid down the basic concepts of modern pacing accurately identifying many of the treatment's problems.
In 1889 he described the application of electricity across the chest to “excite rhythmic contraction… to stimulate by direct means the action of the heart which has been suddenly enfeebled or arrested in diastole by causes of a temporary or transient character”.
Medicine had its first integrated theory of cardiac pacing yet another 20 to 30 years had to elapse before this theory resulted in effective therapy. In particular it had to await significant further medical discoveries (cardiac structure, physiology and conduction pathways) and technical progress (the electrocardiogram, lab stimulators).


The Gerbezius-Morgagni-Adams-Stokes Syndrome


1717

Marcus Gerbazius (1658 - 1718), a Slovenian physician, described the symptoms of bradycardia due to complete atrio-ventricular block.

1761

The Italian Giovanni Battista Morgagni (1682 - 1771) (Fig. 19), founder of pathological anatomy, provided a clinical description of circulatory arrest and implying a causal relationship between a slow pulse and a syncopal attack.

1827

The Irish surgeon Robert Adams (1791 - 1875) described a patient with repeated apoplectic attacks and a slow pulse. He was the first to realise that cerebral symptoms may be caused by cardiac rhythm disorders.

1846

William Stokes (1804 - 1878) (Fig. 20), another Irishman, described once more the pseudo-apoplectic loss of consciousness and bradycardia in a patient, giving further observations on this condition and a detailed analysis of Adam's case.


Atrio-Ventricular Block


1899

Karel Frederik Wenckebach (Fig. 21), a Dutchman, described type I second degree atrio-ventricular block in humans using sphygmographic methods of the radial pulse (the electrocardiogram was not yet in clinical use).

1906

John Hay (Fig. 22) from Liverpool, England, published a case report of type II second degree atrio-ventricular block documenting his findings again without the benefit of electrocardiography. He utilised simultaneous tracings from the radial artery and the jugular venous pulse.

1924

Woldemar Mobitz provided the classification of second degree atrio-ventricular blocks utilising the electrocardiogram, hence documenting Wenckebach's and Hay's findings.


The Electrocardiograph


Over the late 1800's – early 1900's, cardiology witnessed a great technological breakthrough that was to have a major effect on the understanding of arrhythmias and hence on the development of specific therapy including pacing: the invention of the electrocardiograph.


1887

The physiologist Augustus Desire’ Waller (Fig. 23) working in St. Mary's Hospital, London, recorded the first human surface electrocardiogram using the Lippmann capillary electrometer to deflect a light beam (Fig. (Fig.24,24, ,25).25). Alexander Muirhead may have been the first to record a human electrocardiogram but Waller was the first to do so in a clinico-physiologic setting, publishing reports and acquiring extensive experience.
Waller had learnt that “each beat of the heart gives an electric change, beginning at one end of the organ and ending at the other”. He was convinced that he could measure these “electromotive properties of the heart” from the skin surface and proceeded to do so with the electrometer connected between the left and right hands or between the front and back paws of his pet bulldog, Jimmie (Fig. 26).
The clinical significance of the electrocardiogram (Fig. (Fig.27,27, ,28)28) was not recognised at the time and Waller himself said: “I do not imagine that electrocardiography is likely to find any very extensive use in the hospital. It can at most be of rare and occasional use to afford a record of some rare anomaly of cardiac action.”
He would often use Jimmie as the subject when he demonstrated his method at lectures by dipping his legs in pots of saline, which served as the electrodes (Fig.29). A question was raised at the House of Commons and this “cruel procedure” risked being dealt with by the “Cruelty to Animals Act” of 1876. The scientist countered these objections and remarked: “If my honourable friend had ever paddled in the sea, he will appreciate fully the sensation obtained thereby from this simple pleasurable experience.” (Fig. 30) Jimmie never complained anyway! Waller is said to have been quite informal and loved entertaining and dashing around with the newly invented motor car.

1892

Another physiologist, Willem Einthoven (1860 – 1927) (Fig. 31), shares the honour with Waller of having founded this new diagnostic modality. Einthoven recorded the first human electrocardiogram in Europe on April 11th, 1892 using the Lippmann capillary electrometer. He initially indicated the four observed deflections with the characters A, B, C, D but later adopted the middle characters of the alphabeth: P, Q, R, S and T (Fig. 32).
In 1902, he made the first direct recording of the true human electrocardiogram using a modified string galvanometer (Fig. 33).
An extremely thin and light weight quartz “string” was silvered to reflect a beam of light and suspended vertically in a strong magnetic field. The very small electric currents generated by the heart were collected from the arms and legs and conducted to the “string” which was deflected laterally by the passage of this fluctuating current. The “string” threw a a vertical shadow, magnified by a microscope, onto a metal plate in which there is a horizontal slot. This slot allows only a point of the shadow to pass through to a moving photographic plate or film, on which the point of shadow writes in a continuous curve. The photographic material was later developed to produce the image (Fig. (Fig.34,34, ,35,35, ,3636).
The signals were obtained from the two arms and left leg (modern Lead I). To enhance conduction, hands and foot were bathed in saline solution with the tubs wired to the input of the electrocardiograph (Fig.(Fig.37,37, ,38).38). It is interesting to note that the signals were collected from a patient in the University Hospital and transmitted to the physiology lab (quite a good distance away) where the actual recording was made. An early example of telemedicine!
There was wide scepticism by the contemporary scientific community against his methods. But Einthoven continued publishing and in 1913 described the Einthoven triangle as the basis for calculations of electrocardiograms and introduced the bipolar electrode system. Classic rhythms were obtained and published.
Einthoven was formal, methodical and demanded technical perfection. He was keen to apply the modality to clinical problems. In 1924, he was awarded the Nobel Prize for Physiology and Medicine for his electrocardiographic work in developing the string galvanometer.

1933

The extremity bipolar electrode system (the standard electrocardiogram lead system) was expanded in 1933 by F. N. Wilson who introduced the unipolar chest wall electrodes.

1942

E. Goldberger introduced the unipolar amplified (augmented) extremity leads. The 12-lead electrocardiogram as we know it today was now complete!
Several manufacturers to produced commercial versions of the electrocardiograph (Fig. 39). The Cambridge Scientific Instrument Co., headed by Horace Darwin (Charles’ youngest son), produced such a device. The string galvanometer for electrocardiography was superceded by direct writing equipment after the Second World War.
Electrocardiography was of paramount importance in the understanding of cardiac rhythm and hence in the further development of pacing.


Late 1920's – Early 1930's: first pacing machines


Credit for the first external cardiac pacemaker has been shared by two doctors: the Australian anaesthesiologist Mark Lidwell and the American physiologist Albert Hyman. Working independently on opposite sides of the world they developed the first cardiac pacing machines.

1928: Mark Lidwell

Lidwell's device ran on alternating current and required a needle to be inserted into the patient's ventricle. In 1928 he used intermittent electrical stimulation of the heart to save the life of a child born in cardiac arrest. The child apparently recovered completely and survived but not much else is known of Lidwell's efforts. He reported his work to the Third Congress of the Australian Medical Society in 1929.

1932: Albert Hyman

Hyman became interested in reviving the “stopped heart” by means of “intracardial (his term) therapy”. Initially this therapy consisted of the intra-cardiac injection of stimulant drugs such as epinephrine although he soon realised that it was not really the drug that restarted the heart but the needle that set up an action current of injury as it punctured the cardiac wall.
Hyman's device, described in 1932 (Fig.(Fig.40,40, ,41),41), was powered by a spring-wound hand-cranked motor and called by Hyman himself an “artificial pacemaker”, a term still in current use.
The clockwork drove a DC current generator whose electrical impulses were directed into the patient's right atrium through a bipolar needle electrode introduced via an intercostal space (Fig. (Fig.42,42, ,43).43). Pacing could be delivered at rates of 30, 60 or 120 per minute. None of the three models of this device build in the 1930's survives today and only two photos can be traced.
Hyman’ work (Fig. 44) was frustrated and eventually derailed by technical problems and the attitude of the times. The medical and social community was not ready for electrostimulation: Hyman's device was roundly dismissed as “gadgetry” that interfered with natural events at best and the work of the devil at worst. He faced considerable opposition, including that of the Journal of the American Medical Association and did not report his experiments.
No one agreed to manufacture it locally although a battery-operated version (lost to history) was eventually manufactured by Siemens-Halske in Germany and their American subsidiary Adlanco (Fig. 45). The Hymanotor, as it was called, was tested but found ineffective and again unfavourably reported upon. During the Second World War, Hyman unsuccessfully urged the US Navy to support his device for use in resuscitation of dying servicemen.
It is important to note that Hyman intended the pacemaker to restore a normal heartbeat in patients whose heart had stopped accidentally or in stillborn infants rather than in those with heart block. In the mid-1930's, the connection between Stokes-Adams disease and pacing had not yet been made.


Early 1950's: first mains-powered portable pacemaker


Mains-powered pacemakers were developed in the early 1950's and were large bulky boxes filled with vacuum tubes that could not of course be implanted. They had to be wheeled around on carts and plugged into wall mains socket outlets to obtain their alternating current power. They were portable only in name since they could only go as far as the nearest electrical outlet!

1949

In Toronto, Canada, Wilfred Bigelow (Fig. 46) and John Callaghan started using hypothermia to reduce metabolism and produce bradycardia and asystole to permit cardiac surgery. Re-warming could not however restore cardiac contraction sufficiently rapidly and so the surgeons started experiments with sino-atrial node stimulation.
In 1949, during an experimental operation on a dog, the heart suddenly stopped. Bigelow recounts: “Out of interest and in desperation, I gave the left ventricle a good poke with a probe I was holding. All four chambers of the heart responded. Further pokes clearly indicated that the heart was beating normally with good blood pressure.” The electric pacemaker was developed as a direct result of these hypothermia experiments.
During the 1940's and early 50's the principle device available to generate a variety of electrical impulses, potentially capable of stimulating the heart was a physiological stimulator manufactured by Grass Manufacturing Co for clinical and physiology lab application (Fig. 47). It used a thyratron rectifier tube to convert alternating current into direct current suitable for stimulation of biologic tissue. The stimulation rate, voltage output and pulse width could be varied (monophasic rectangular pulse of 2-20 ms duration).
John Hopps (Fig. 48), an electrical engineer, was recruited on a part-time basis by the National Research Council of Canada and designed what was perhaps the first electronic device specifically built as a cardiac pacemaker. It was an external unit driven by vacuum tubes. The electrical impulses were transmitted via a bipolar catheter electrode to the atria using a transvenous approach. Atrial pacing was readily achieved and heart rate could be controlled with no uncomfortable chest wall contractions (Fig. 49).

1951

Paul Zoll (Fig. 50), a Boston cardiologist, is given credit for ushering in the modern era of clinical cardiac pacing. He had read the work done by Callaghan, Bigelow and Hopps and developed an external tabletop pacemaker that was successfully applied to the treatment of heart block (Fig. 51).
The electrodyne PM-65 pacemaker, designed by Zoll, comprised an electrocardiograph to monitor the cardiac rhythm and an electric pulse generator to pace the heart. The pulse generator was a modification of the electric stimulator used in physiology laboratories. It delivered periodic electric impulses at 2 ms pulse width and 50 to 150 volts alternating current pulse amplitude through a pair of 3 cm2 metal electrodes strapped to the patient's chest directly over the heart. The electrodes irritated the skin and the patients of course found the repeated electric shocks painful.
The mains-powered unit was bulky and heavy and was carried on a cart. It could only go as far as the extension cord would allow (Fig. 52).
In 1952, he reported on two patients suffering from recurring prolonged ventricular standstill whom he treated with this external device (Fig. 53) and in 1956, he applied transthoracic electric shocks to reverse ventricular fibrillation in humans and soon after developed the first cardiac monitors for clinical use.

1956

In St. George's Hospital, London, Aubrey Leatham (Fig. 54) and Geoffrey Davies developed an external stimulator with which to resuscitate patients with heart block and asystole. The first studies on the use of external pacing for cardiac standstill had just been reported by Paul Zoll in Boston; however, Zoll's pacemaker was a fixed system without a demand capability and could cause an “R on T”-induced ventricular fibrillation. At St. George's, Leatham asked Davies to develop the first demand circuit device which was published in 1956.
This mains-powered device stimulated the heart through the intact chest wall utilising 150 volts and was commercially manufactured by Firth-Cleveland in the UK. The commercial version contained several modifications: duration of asystole permitted, sensitivity controls to sense the electrocardiogram, two output ranges and a battery for independent operation.



Late 1950's – early 1960's: The “Golden Years”


These years witnessed several important achievements in the field of cardiac pacing by multiple persons and their teams working in different parts of the world: they were the “golden years” of pacing.
Three landmark “firsts” will be described in detail: the first battery-operated wearable pacemaker (1957), the first totally implantable pacemaker (1958) and the first long-term correction of heart block with a self-contained, implantable pacemaker (1960). These events had far-reaching consequences and opened up the field to the future.


1957: First battery-operated wearable pacemaker


Earl E. Bakken, electrical engineer, TV repairman and co-founder of Medtronic Inc. produced the first battery-operated wearable pacemaker.


The engineer

Earl E. Bakken (Fig. 55) and his brother-in-law Palmer Hermundslie had co-founded Medtronic on April 29th, 1949 in a garage in northeast Minneapolis (Fig. (Fig.56,56, ,57).57). The company had led a precarious existence as a repair service for hospital electrical equipment and regional distributor for other manufacturers. They would build new equipment on order or customise standard instruments for laboratory or clinical researchers. They would hang around hospital surgical suites setting up equipment, training personnel in its use and troubleshooting and repairing it as necessary. Meanwhile they forged working relationships with physicians and their staff.
Bakken recounts: “We never made any serious money on that early custom-building activity, rarely even recouping the cost of the prototype… everything we did, we lost money on!” They were however present at the right time in the right place!


The surgeon

C. Walton Lillehei (Fig. 58) was a leading cardiac surgeon at the University of Minnesota, Minneapolis and had attained international fame by the mid-50's. Techniques had been developed to enter the heart and correct congenital defects while the circulation was supported. By 1957, Lillehei had performed over 300 open-heart operations on young adults and children. This rapidly evolving field of open heart surgery was to be a major driving force towards the development of cardiac pacing.


A series of problems… resolved!

Despite successful repair of the congenital defect, about 1 patient in 10 developed post-operative complete heart block due to damage of the conducting system while the surgical repair was being performed. Stimulant drugs such as adrenaline, atropine, or the newly developed isoprenaline, were helpful in the short-term but proved disappointing over a longer time frame and could not prevent sudden recurrence of heart block. Another solution had to be found!
It was thought that temporary rhythm support via pacing would keep the patient alive until recovery of the conducting system occurred. The technology developed by Zoll was clearly inappropriate as the high voltage pacing stimuli delivered trans-thoracically would be far too traumatic on these young children.
The physiologist John Johnson proposed the utilisation of the Grass stimulator that was used in the physiology labs to activate hearts. After several experiments, Vincent Gott and William Weirich concluded that a cardiac rhythm could be restored in animal hearts in which heart block had been surgically created by means of a wire inserted into the wall of the right ventricle and connected to the external stimulator. Low voltage pulses at the desired rate could easily stimulate these hearts.
Lillehei and his co-workers developed the myocardial wire: a multi-stranded, braided stainless steel wire in a Teflon sleeve (Fig. 59). One end of this was implanted directly into the myocardium and the other end was exteriorised via a stab incision and connected to the physiology lab stimulator. An indifferent electrode was buried under the skin to complete the circuit. Effective pacing needed only 1.5 volts as there was direct contact with the myocardium. There was no rejection and no damage to the beating heart and the wire could be removed easily by tugging once normal conduction resumed.
The first myocardial wire was implanted on the 30th January 1957 in a 3-year old girl in whom heart block had complicated the repair of Fallot's tetralogy. Pacing was successful and the little girl soon regained sinus rhythm and survived. Myocardial wires started being implanted electively, ready for immediate use later should this become necessary. A technique for their implantation through a hollow needle was also developed for non-surgical patients who developed Stokes-Adams attacks.
Further problems soon became obvious: the stimulator was large and heavy, of limited portability and awe-inspiring especially for paediatric patients. Moreover, the system was fatally flawed since it depended totally on its external mains power supply and on the length and integrity of the extension power cord. If power supply failed, it was worthless.
On October 31st, 1957 a municipal power failure lasting three hours resulted in the tragic death of a baby. The hospital had emergency power generation in its surgical suites and recovery area but not in its patient rooms. The caregivers were once more reminded of the limitations of existing technology.
The day after, Lillehei requested Bakken to see if Medtronic could come up with something better. Patient mobility needed improvement and concerns about power failure needed to be eliminated. When Bakken accepted Lillehei's assignment, it seemed to him just like any other special order for a piece of equipment. This was not to be!


First attempts

Initial attempts at building a more reliable and portable pacemaker involved adding an automobile battery with an inverter to convert 6 volts direct current into 115 volts alternating current and then power the conventional alternating current pacemaker on its wheeled stand.
These plans were soon abandoned, as they were obviously highly inefficient! A 10 volt direct current pulse was sufficient to stimulate the heart and transistors were becoming widely available.


The prototype

Bakken dug out the April 1956 back issue of Popular Electronics in which he recalled seeing a circuit for an electronic, transistorised metronome. The circuit transmitted clicks through a loudspeaker: the rate of the clicks could be adjusted to fit the music. The blocking oscillator circuit that was utilised had actually been invented at the MIT Radiation Laboratory during World War II.
He simply modified the two-transistor circuit (Fig. 60) and placed it, without the loudspeaker, into a four-inch-square and inch-and-a-half-thick aluminium box with terminals and switches on the outside. The circuit was powered by a powerful miniature 9.4 volt mercury battery housed within the box. There was an on-off switch and control knobs for stimulus rate and amplitude (Fig. 61).
Bakken recounts: “Without any grandiose expectations for the device, I was moderately optimistic about what it might eventually do for Lillehei's patients. I drove the device over to the University's animal lab where it could be tested on a dog. Of course it worked.
“The next day I returned to the hospital to work on another project when I happened to walk past a recovery room and spotted one of Lillehei's patients. I must have done a double take when I glanced through the door. The little girl was wearing the prototype I had delivered only the day before! I was stunned. I quickly tracked down Lillehei and asked him what was going on. In his typical calm, measured, no-nonsense fashion he explained that he’d been told by the lab the pacemaker worked and he didn’t want to waste another minute without it. He said he wouldn’t allow a child to die because we hadn’t used the best technology available.”
After only 4 weeks of experimentation and work, the first battery-powered, transistorised pacemaker was already in clinical use! A feat that is unlikely ever to be repeated given the regulatory labyrinth that all devices have to go through from inception to clinical use.


The “first ten”

The first production run of ten or so units were more refined versions of the original prototype and went into clinical use soon after at the University (Fig. 62). The dials had been recessed so that children would be less likely to adjust them and a little neon light blinked red with each stimulus. In addition, two metal handles (borrowed from an old ECG machine) were been added such that a strap could secure the pacemaker to the body. The pacemaker was not only portable but wearable! This pacemaker became known as the 5800 (because it was made in 1958). The product literature (Fig. 63) stated boldly:
“So small and light that it may be attached to and worn by the patient, the Medtronic Cardiac Pacemaker stimulates ventricular function in cases of atrio-ventricular dissociation that are induced during the surgical repair of septal defects, or that occur spontaneously as in Stokes-Adams syndrome. The Pacemaker is designed for internal applications with at least one wire attached directly to the myocardium for temporary stimulation or with a bipolar patch for prolonged stimulation.
“Created with imagination and originality, the transistorised circuit completely removes the hazards and nuisance associated with AC powered instruments. Its self-contained miniature power source will operate the instrument for approximately 1000 hours.”
The chosen pacemaker output was a 2 ms square wave, variable in amplitude from 1 to 20 mA into a 1000 Ω load. The blocking oscillator repetition rate was variable from 60 to 180 pulses per minute. The box was “wearable” (Fig. 64).


The significance and the consequences

Medical historians regard Bakken's pacemaker is one of the first successful applications of transistor technology to medical devices helping to launch the new field of “medical electronics” (Fig.65). In the entire history of medicine before 1957, there had never been a partly or completely implantable electrical device. It was however apparent that for long-term pacing a totally implanted device would have to be designed as ascending infection via the pacing electrodes occurred frequently.
Lillehei himself noted: “The question of how long stimulation can be maintained appears to be related to electrode materials, design and technique of implantation… The possibility of infection along the wire exists, but… can be minimised by tunneling the wire for some distance before bringing it out on through the skin.” (Fig. 66) Most patients with post-operative heart block regained sinus rhythm within a few weeks but one patient was kept on the device for 15 months.
Recurrent heart block in patients who had recovered from their post-operative heart block caused several deaths. It was apparent that these patients needed indefinite and not temporary pacing for them to survive. The myocardial wire developed exit block as scar tissue grew around the site of stimulation increasing electrical resistance and requiring a progressive increase in pacing stimulus voltage to maintain capture. The thoracic muscles began to twitch at these increased voltages. A totally implantable system with better designed elctrodes neede to be designed!
Meanwhile elsewhere, on the 16th July 1958 a transvenous catheter electrode was introduced fluoroscopically, via the basilic vein into the right ventricular outflow tract, in a patient with fixed complete heart block who required colon resection because of a malignancy. Pacing was continued for two hours, during the operative procedure, and ended with slowing of the stimulation rate until an unpaced idioventricular rhythm developed. The catheter was removed without complication and the patient resumed the idiventricular bradycardia.


1958: First implantable pacemaker

On October 8th, 1958 the first pacemaker implantation was performed in Sweden. The system had been developed by the surgeon Ake Senning and the physician inventor Rune Elmqvist and implanted on a 43-year old engineer called Arne Larsson. This first experience with a fully implantable pacemaker system was reported at the Second International Conference on Medical electronics in 1959 and published as an abstract in 1960 (Fig. 67).
The patient suffered from Stokes-Adams attacks that required resuscitation many times daily and whose situation was considered hopeless. The implantation was a more or less desperate rescue measure. The risks taken with this completely unknown therapy were immense.


The scientists

Ake Senning (Fig. 68) was the cardiac surgeon in charge of the Department of Thoracic Surgery at the Karolinska Hospital in Stockholm. He had observed Lillehei's work with temporary external pacing.
Rune Elmqvist (Fig. 69) was a medical graduate who had not pursued a medical practice but became an engineer. He had designed a portable ECG machine in 1931 and then the widely used ink jet recorder, the Mingograf, in 1948.
These two men began to collaborate closely in 1950 and developed fibrillators and defibrillators for open heart surgery. They realised that the main problem with external pacemakers was the open route for ascending infection along the lead and decided to design a fully implantable system.


The patient

Arne Larsson (Fig. 70) is the first human to receive an implanted pacemaker. He had been hospitalised with complete heart block and frequent Stokes-Adams attacks for 6 months. He was having 20 to 30 attacks daily and his prognosis was poor. Treatment was maximised with ephedrine, pentymal, atropine, isoprenaline, caffeine, digoxin and whisky.


The woman: Else Marie Larsson

Else Marie was the patient's wife who pleaded with Elmqvist and Senning to help her hopelessly ill husband. She had read press reports about ongoing experiments with electrical stimulation of the heart and hounded down the two scientists for a solution that did not yet exist: an implantable pacemaker.
Senning recounts his encounter with this lady: "An energetic, beautiful woman entered my lab on the 6th October 1958 and told me that I had to implant a pacemaker into her husband. I told her we had not completed our experimental series and we did not have a pacemaker for human clinical implantation. She demanded: ‘So make one!’. That day she drove several times from Elmquist's electronic lab and back and finally convinced us.”


The procedure

To avoid publicity, the implantation was done in the evening when the operating rooms were empty. Via a left-sided thoracotomy two suture electrodes were implanted into the myocardium and tunnelled to the pacemaker box placed in the abdominal wall. The first pacemaker implanted functioned only a few hours but the second one implanted in the same patient had better longevity.
Senning recounts: “On the 8th October 1958, in the evening, when there were no extra people in the theatre, I implanted the first pacemaker, but it lasted only 8 hours. Presumably, I had damaged the output transistor or capacitance with the catheter and I did not have the other one which was in the lab. I implanted the other one early the next morning”. Senning then concludes: “In the 1950's we did not have any liability problems. The patient and relatives were happy if the patient survived.”
The second pacemaker functioned well for about 1 week before suddenly showing a decrease in the ECG pacing stimulus size: suggesting probable lead fracture rather than pulse generator malfunction.


The circuit

The pulse generator delivered impulses at an amplitude of 2 volts and a pulse width of 1.5 ms. The pulse rate was fixed at a constant rate of 70 to 80 beats per minute. The energy utilised was minimised since Elmqvist managed to obtain a few of the first silicon transistors imported into Sweden. These were more efficient than the older germanium transistors. With them Elmqvist designed a stable and efficient blocking oscillator with a small power consumption (Fig. 71).
The first transistor forms a repetitive blocking oscillator whose pulses are fed to the base of the second transistor. the collector of this second transistor is then connected to the pacing electrode over an RC network.
Several types of primary battery cells could have been used. The Ruben-Mallory cells with zinc as the anode and mercuric-oxide as the depolarizer were a possible choice (Fig. 72). They had been invented during World War II for army field telephones. Although the cell potential remained constant, these cells had a short lifetime and released hydrogen gas at the zinc anode. The effect of this gas in a cell encapsulated in plastic was not known. For these reasons, nickel-cadmium rechargeable cells were then chosen. Two cells of 60 mAh each were sealed, encapsulated and connected in series.
Recharging was accomplished inductively. A coil antenna with a diameter of about 50 mm was connected to the cells via a silicon diode. This was inductively coupled across the patient's skin to a large external flexible coil 25 cm in diameter attached to the patient's abdomen with adhesive tape. Recharging was accomplished by a 150 kHz radio frequency current generated by an external mains-powered vacuum tube device connected to the external coil. The pacemaker required charging once a week for 12 hours.


The device

The entire unit was entirely hand-made (Fig. 73) and consisted of the nickel-cadmium batteries, the electronic circuit and the coil recharging antenna. These were encapsulated in a new epoxy resin (Araldite) produced by Ciba-Geigy, which had excellent biocompatibility. The approximate diameter and thickness became 55 mm and 16 mm respectively, according to the dimensions of the ever so popular shoe polish can from Kiwi (Fig. 74). Elmquist in fact produced two such units using these cans as moulds!
These first units had two electrode wires, each consisting of a twinned, stainless steel suture wire with polyethylene insulation. The distal ends of the wires were sewn into the myocardium to act as pacing electrodes. The proximal ends were hard-wired to the pulse generator circuit. It was estimated that the electrode had to withstand about 105 bends per day (Fig. (Fig.75,75, ,7676).


The personal outcome

Rune Elmqvist soon ceased his involvement in pacing but remained active in other areas of medical technology. He died in 1997, aged 90. Ake Senning remained very active in the field of cardiac surgery. He died in 2000 at the age of 84. Arne Larsson survived both the engineer as well as the surgeon who had saved his life (Fig. 77). He required five lead systems and 22 pulse generators of 11 different models until his death on December 28th 2001 aged 86 of a malignancy totally unrelated to his conduction tissue disease or his pacemaker system.


1960: The first long-term correction of complete heart block


Wilson Greatbatch

Wilson Greatbatch was an electrical engineer teaching at the University of Buffalo where he was working on an oscillator to aid in the recording of tachycardias. He accidentally discovered the way to make an implantable pacemaker (Fig. 78).
Greatbatch, a deeply religious man, describes the event this way: “It was no accident, the Lord was working through me… The oscillator required a 10 KΩ resistor at the transistor base. I reached into my resistor box for one, but I misread the color coding and got a 1 MΩ resistor by mistake.”
When he plugged in the resistor, the circuit started to “squeg” with a 1.8 millisecond pulse followed by a 1 second interval during which the transistor was cut off and drew practically no current. “I stared at the thing in disbelief,” he said. Wilson Greatbatch immediately realized that this small device could drive a human heart but it wasn’t easy to find a heart surgeon who would believe in his idea.


Dr. William Chardack

Dr. William Chardack was chief of surgery at Buffalo's Veteran's Hospital at the time. In Dr. Chardack, Greatbatch had finally found a surgeon who believed in the viability of an implantable pacemaker.
On May 7, 1958, Greatbatch brought what would become the world's first implantable pacemaker to the animal lab at the hospital. There, Chardack and another surgeon, Dr. Andrew Gage, exposed the heart of a dog to which they touched the two pacing wires. The heart proceeded to beat in synchrony with the device. The three looked at each other. Their feelings were best expressed by Dr. Chardack, who exclaimed, “Well, I’ll be damned.”
In a lab book about a year later, Greatbatch wrote “I seriously doubt if anything I ever do will give me the elation I felt that day when a 2 cubic inch electronic device of my own design controlled a living heart.”

The “bow tie team” (Fig. 79)

The three - Greatbatch and Drs. Chardack and Gage - became known as the bow tie team. “The two doctors wore bow ties because children tend to pull long ties. I wear bow ties because long ties get in the way when I am soldering.” The months and years that followed involved a great deal of research and experimentation. “I frequently took problems to the Lord in prayer,” Greatbatch says, “and I always got the answer.”
Over the first two years experiments were made with animals. In 1959, Greatbatch patented the implantable pacemaker, and William Chardack reported the first success in a human with this unit in 1960. The procedure was completed in June 1960 on a 77-year old man in complete heart block (Fig. 80). Chardack first implanted the lead and when threshold stabilised implanted the pulse generator. The patient survived uneventfully for 2 years before his death from natural causes.
In 1961, Chardack, Gage and Greatbatch reported a series of 15 patients who had pacemakers implanted. Greatbatch later invented the long-life corrosion-free lithium-iodine battery to power the pacemaker (Fig. 81).

The global outcome

The early pacing technology of the 1950's and 1960's was a spin-off from the research and development of World War II and Cold War eras.
Faulty batteries, body fluids leaking into the encasement and broken leads caused numerous pacemaker failures that required emergency surgery. The main difficulty however was the lead. It was soon obvious that the myocardial wire was unsuitable as a long-term electrode. Stimulation threshold increased after a few weeks until exit block developed and no more capture was possible. Moreover, the wire could not resist the enormous repetitive mechanical stresses of bending. These technical problems contributed to the delay in the widespread use of implanted pacemakers for several years.
Tight collaboration between engineers, physicians and patients was the fundamental driving force for the growth of a significant global industry. Well over 2 million pacemakers have been implanted worldwide since 1960!
  • Zoll founded Electrodyne and continued developing pacemakers.
  • Earl Bakken (co-founder of Medtronic Inc.) started producing the Chardiack-Greatbatch pacemaker.
  • Wilson Greatbatch, after a time with Medtronic, founded his company (Wilson Greatbatch Ltd.) and convinced the industry to change from mercury to lithium-iodine cells.
  • The company Elema Schonander, for which Rune Elmqvist worked, became Siemens-Elema in 1974. Siemens then acquired Pacesetter Inc. in 1985 and combined them to form Siemens-Pacesetter which was then in turn acquired by St. Jude Medical in 1994

1958-1959: Inductively coupled cardiac pacemakers


Other investigators followed a different line of approach in designing self-contained implantable pacemakers: inductive coupling (Fig. 82).
A pair of electrodes were sutured to the epicardium and connected to a coil antenna located subcutaneously. Minimal or no circuitry was implanted and no internal batteries were needed.
This coil antenna was inductively coupled to an external coil taped to the patient's intact skin. This external coil was connected in turn to a transistorised pulse generator powered by an external battery. The electronic components, relatively unreliable at this time, were therefore located entirely outside the body.
Glenn, Mauro, Longo, Lavietes and Mackay's technique utilised a radio-frequency oscillator . Later versions of this system included triple-helix, silicone insulated endocardial leads and rate-control via an external knob (which the patient himself could modify at will). Atrial pacing with this device was used in 1969.
Inductively-coupled pacemakers proved to be very successful with several hundreds of implants and survival rates of over 10 years (Fig. 83). These devices were extensively used in the Birmingham (UK) region for a number of years, being produced by the Lucas factory, more commonly known for its automotive electrical products (until taken over by Bosch). One particular disadvantage of this device was that its removal (for example, for bathing) could result in bradycardia and syncope. They continued to be used until well into the 1970's and several patients with later generation pacemakers still have the implanted coils from their original devices.


1959: Hunter – Roth electrode

On the 4th April 1959 Samuel Hunter (Professor of Surgery at St. Paul) and Norman Roth (Chief Engineer at Medtronic) implanted a bipolar stainless steel electrode to pace a patient suffering from post-myocardial infarction complete heart block (Fig. 84). The lead consisted of a pair of stainless steel wires secured in a silicone rubber base (Fig. 85).


1959: Elema-Ericsson lead

A new lead was developed in 1959 by Elema Schonander and the Telecom Company, Ericcson. This consisted of four thin bands of stainless steel wound around a core of polyester braid and insulated with soft polyethylene (Fig. 86). It was estimated to resist over 184 million flex cycles, hence lasting for at least 6 years. The unipolar epicardial stimulation electrode was a platinium disc, 8mm in diameter and insulated at the back.

1959 - 1960: Elema Schonander

The Elema 135 (Fig. 87) rechargeable pacemaker was successfully implanted in Stockholm (1959), Uruguay (February 1960) and England (March 1960) but Elema Schonander never filed a patent application. The maket prospects were perceived to be poor! Pacemakers were considered as an expensive service to prominent customers with little commercial value. The external charging system was too complicated especially for elderly patients.
Elmqvist constructed the Elema 137 pacemaker in 1960 (Fig. 88). Ruben-Mallory zinc-mercury oxide cells were used as the power source thus eliminating the need for periodic recharging of the previously utilised nickel-cadmium cells.


Early 1960's

Other models were implanted with similar success in 1961 by Zoll et al (Fig. 89) and in 1962 by Kantrowitz et al. The technique for inserting permanent transvenous bipolar pacing electrodes was developed in 1962 by Parsonnet et al. (in the US) and by Ekstrom et al. (in Sweden).

Recent history

Pacemaker and lead technology continued to develop rapidly to make these devices reliable, automatic and flexible in the therapy they provide. The therapeutic end-point shifted from saving life to enhancing its quality and simplifying follow-up. Electrotherapy has become socially accepted and its indications are extending also to non-cardiac pathology: Parkinson's Disease, pain-control, drug delivery.

Mid 1960's (Fig. 90)

Transvenous leads replaced epicardial leads. Pacemakers and their leads could be implanted without a thoracotomy and without general anaesthesia. “Demand” pacemakers were developed to sense the underlying cardiac activity and provide pacing only when needed.

1970's

Lead design improved: “tined” for passive fixation and “screw-in” for active fixation. The lithium-iodine battery was developed to replace the mercury oxide-zinc battery that had been used till then. This resulted in greatly increased pacemaker longevity (Figs. (Figs.9191 and and9292).
In 1972 an American-made radioisotope pacemaker was implanted by Parsonnet et al. These nuclear pacemakers had an expected life of 20 years but went out of fashion mainly due to the need for extensive regulatory paperwork (Fig. 93).
Titanium casing was developed to enclose the battery and circuitry. This replaced the epoxy resin and silicone rubber that was previously utilised to encase the internal components of the pacemaker.
Pacemakers were made non-invasively programmable in the mid-1970's. Using a radio-frequency telemetry link, most pacing parameters could be adjusted to follow the changing clinical needs of the patient.
By the end of the 70's dual-chamber pacemakers were developed to pace and sense in both atria and ventricles. Synchronised timing made it possible to preserve the atrial contrbution to ventricualar filling as well as to track the intrinsic atrial rate.

1980's

In the early 1980's steroid-eluting leads were developed. These eluted steroid from their tip and hence decreased the inflammatory response evoked by the presence of the lead tip (acting as a foreign body). Consequently, the early rise of capture threshold was blunted and safety was enhanced (Fig. 94).
In 1981, Zoll patented and re-introduced a transcutaneous external pacemaker with a longer pulse width of 40 ms and a larger electrode surface area of 80 cm2. This reduced the current necessary to capture the heart and thus improved patient comfort. This method of pacing could be applied very rapidly as a bridge to a the establishment of pacing via the transvenous route.
In the mid-1980's rate-responsive pacemakers were designed. A tiny sensor within the pacemaker box detected body movement and used this as a surrugate measure of activity. Signals from the sensor were filtered and applied to an algorithm to alter the pacing rate up or down. Thus, pacing rate would change according to the patient's activity level.

1990's

Microprocessor-driven pacemakers appeared. These became very complex devices capable of detecting and storing events utilising several algorithms. They delivered therapy and modified their internal pacing parameters according to the changing needs of the patient in an automatic manner. The rate-response pattern also adjusted itself automatically to the patient's activity level (Fig. 95).

2000's

Bi-ventricular pacing for heart failure was introduced. An additional specially-designed lead was introduced via the coronary sinus to the epicardial surface of the left ventricle. The right ventricle (via the standard lead) and the left ventricle were paced simultaneously to attempt to resynchronise contraction of the left ventricular septum and left ventricular lateral walls. The improved contraction improved symptoms and survival (fig. 96).
Automaticity progressively increased thus making follow-up visits easier and briefer. Pacemakers could also upload data telephonically to a central server via the internet (Fig. 97).

Conclusion

The history of pacing (Fig. 98) is an exciting story of initiative and innovation, often in the face of criticism and opposition. It is a unique mix of medicine, technology and marketing which has developed into a major industry and has brought electrotherapy out of the labs and into the clinics.
Perhaps the single most important event that enabled the development of this form of therapy was the invention of the transistor in December 1947 (Fig. 99). Indeed one of the first applications of the newly invented device was in the nascent field of medical electronics, and particularly in pacemakers.

Suggested reading

One man's full life by Earl Bakken
Landmarks in Cardiac Surgery by Stephen Westaby
First Artificial Pacemaker: a milestone in the history of cardiac electrostimulation http://asmj.netfirms.com/article0903.html
The Bakken; a library and museum of electricity in life http://www.thebakken.org/
Earl Bakken's Little White Box: the complex meanings of the first transistorised pacemaker by David Rhees and Kirk Jeffrey
Wilson Greatbatch: man of the millennium by Joseph Radder
History of electrotherapy http://www.hrsonline.org/ep-history/
History of Cardiac Rhythm Disorders by B Luderitz
A Brief History of Cardiac Pacing by Glen Nelson

Credit to Oscar Aquilina, Consultant Cardiologist and Electrophysiologist, Department of Cardiology, St. Luke's Hospital, Guardamangia, Malta.

Topics

SMART-WPW

  A Novel ECG Algorithm for Accurate Localization of Manifest Accessory Pathways in Both Children and Adults: SMART-WPW The SMART-WPW algori...